SEM: The SCANNING ELECTRON MICROSCOPE and Scanning Electron
Microscopy
This document gives an overview of the basic operation of the scanning
electron microscopy (SEM) and in particular, the
Cryo-SEM (cryogenic scanning electron microscope).
The SEM we use is the JEOL JSM-6400, which is in the MicroLab, Department of
Chemical Engg., IIT-B, Powai, Mumbai - 400 076. All users of this
microscope are required to know the information given in this
document. Your comments are welcome. For further
information, please contact Prof. Jayesh Bellare.
Email: jb@iitb.ac.in
BRIEF DESCRIPTION OF APPARATUS
The JSM-6400 basic unit consist of an electron optical column
mounted on a main console, a control and display system, a power supply
unit and a pump box. The main console incorporates a vacuum system and
the control and display system, photographic recording system (camera,
film holder etc.) power supply unit is placed at the back of the
control and display system.
PRINCIPLE
When a specimen is targeted with primary electron beam, the
complex interactions between the specimen and the electrons give rise
to a variety of signals that are detected in
the SEM. The SEM derives information from scattered electrons and
other signals like photons as they are reflected from the surface of the bulk samples and
transmitted into the samples. Back scattered electrons are those
that have undergone single or multiple scattering events in the sample
and have escaped through the surface of the specimen. BSE travels in
straight lines and only those electrons directly in line with detector
will be used to form the image. This gives rise to useful shadowing
effects particularly on fractured faces. Secondary electrons are
transmitted by very low energy and are emitted from the first 5-10 nm
of specimen. This shows fine structure details. A substantial
proportion of electrons are transmitted if sample is thin enough. These
transmitted electrons provide a mean of examining the internal
structure of sample in a transmission electron microscope (TEM), but
not used in an SEM. Cathodoluminescence is the phenomenon where
emitted photons are collected and analysed.
The JSM-6400 is associated with modern digital image
processing system. This system enables image enhancement by averaging
and integration and sorting of image data in frame memories. It can
accommodate optical attachment for cathodoluminescence, X-rays,
absorbed electrons, transmitted electrons.
APPLICATIONS:
Microscopy is a useful tool to directly get images of a sample at
high magnification so that features smaller than what the unaided eye
can see are revealed. SEM uses electrons so as to give high
resolution and high depth of field. However, samples must be
compatible with high vacuum and with the electron beam. Samples are
made compatible by a process of "sample preparation." Cryo-SEM is used in special cases where usual sample preparation fails to give correct results.
Cryogenic electron microscopy can be used to
- gain microstructural and morphological information of various
systems that contain liquids or are of high vapour pressure, e.g.,
soft solids, comples liquids like liposomes
and vesicles.
- study the dynamics of microprocesses by capturing instants of an
evolving process
- act as a microanalyser (of electron-induced X-rays). By using an energy
dispersive spectrometer allowing an accurate and efficient
nondestructive elemental analysis and elemental distribution of
consitituents.
These capabilities are very useful in fields of material science,
biology and industrial research. It is becoming increasingly
important in the study of chemical processes and materials, and hence
is a very useful tool for chemical engineers and materials scientists
to master.
Important Instructions for the use of SEM
All users of the MicroLab must
follow these instructions scrictly when using the SEM:
-
Check the logbook first for any entries in the "remarks" column, and
only then turn on the SEM.
-
Follow the prescribed steps in the correct order to turn on the SEM
-
When turning the SEM ON, put on the A/C and the dehumidifier 15 minutes
(-0/+5 minutes) before turning on the chiller.
-
Keep the A/C, dehumidifier and chiller ON throuout your session, i.e.,
as long as the SEM is ON. Due to the low room temperature, you might feel cold during the session, so
always keep a jacket or sweater handy.
-
Do not run a sample until you have the In-Charge's explict written
permission. There are many compatibility issues that may spoil the
instrument, so never run a sample without permission. There are also
some legal requirements that we all have to comply with.
-
While we welcome visitors, it is only by appointment that they can see
the instrument. Do not get your friends or any "tourists" to witness the instrument
while you are operating. Crowds disrupt the clean environment and
temperature. Special demo sessions can be organised for
them. Ask the In-Charge for details.
-
Do not take out anything from the lab. Specifically, do not take out
the stubs, stub holder/cryo-stage, etc.
-
Due to the nature of the entire MicroLab room environment, you are
advised not to
have any food or drink there. Such activity may take place in your
other sitting area, e.g., Res.Schol room or Silicate lab (outer room).
-
When turning things OFF, put off the chiller 10 minutes (-0/+2
minutes) after turning the SEM OFF, then wait an additional 20 minutes
(-0/+5 minutes) and then turn OFF the A/C and the dehumidifier.
-
Do not keep any tools/samples inside the SEM room. The only things
you should have to take inside the SEM room are the prepared
specimens.
-
At the end of the session (perhaps whilst waiting for the prescribed
time limits to pass), clean up all the working areas completey.
-
All accessories of the sample preparation room must be kept only in
the sample preparation room in the large thermocole box. Arrange the
items neatly in the box. Keep all personal/individually issued items
(like rolls, gold wire, etc.) in your locker.
-
Clean up completely after yourself: paper, samples, shoes, furniture,
bits of film, etc. Ensure there are no items strewn
about. Arrange all furniture neatly.
-
Keep all lN2 and freon cryocans and cylinders in designated
positions and locations.
-
Take particular care in ensuring that the darkroom is clean and dry;
wipe the platform dry before you leave. Ensure that spirals are
washed, cleaned and dried.
-
Help in cleaning up the lab: dust/mop/vacuum once a week or more
frequently.
-
If the lab is dirty when you enter it, it is your duty to note exactly
what was wrong in the logbook in the remarks column. Clean up
whatever you can, and claim restitution from the previous user.
Otherwise, you
will be held responsibe for it.
-
Before you leave, ensure that all things are in their default
locations and settings.
-
Default settings of the SEM are:
-
Stage position 35, 25, 000, 222
-
Accn. voltage 5 kV
-
Filament emission to zero
-
Probe current to extreme right (16)
-
Remove samples from the stage
-
Left and right screen setting reduced using knob below the screen
-
Mag to 12x
-
Camera film to be removed; if exposures are left, forward to next
exposure and leave a note on the camera.
-
Insert camera slider plate.
-
Use PF1 to blank right screen
-
Ensure that you know the purpose and function of each piece of
equipment or tool that you use. You are responsible for this.
-
Sample and roll details shoulr be maintained in the format provided in
the MicroLab and also online.
Various procedures for the SEM are given below.
The "Switch-On" Procedure is:
- Check the SEM logbook for any relevant comments
- Check the entire MicroLab for any irregularities like missing
things or dirty areas.
- Switch on both the A.C.s in the microscope room.
- Switch on the cooler and ensure that it turns on correctly by noting
that you can hear the cooler's
compressor sound.
- Wait for about half an hour.
- Switch on the water circulation pump.
- Open the Nitrogen gas cylinder valve and adjust/check that the pressure is 5
kg/cm2.
- Switch on the "To Stabilizer" switch in the sample preparation
room.
- Check the voltage to be at 230 volts. If not at 230 volts, adjust
to 230 volts by changing to manual mode and put it back to auto mode.
- Switch on the stabilized supply.
- Switch off one of the A.C. s in the microscope room.
- Turn the "Car ignition" key to on position. Wait for five seconds
and then to start position. Hold it for 3 - 4 seconds and take your
hand off.
- Wait until ready signal comes on (approximately 40 minutes).
- The specimen is then imaged. (See below "IMAGING THE SPECIMEN")
- Note: To the extent possible, the microscope should be kept
continuously on. Turn it off if you are the last operator of the day.
Frost was found to be minimum when the microscope is
continuously kept on. Samples can be made even before the microscope
is switched on. They can be stored under liquid nitrogen until needed.
Imaging The Specimen
- The accelerating voltage is set at the desired value (generally
between 5 KV to 10 KV for cryo specimens) and is switched on.
- The Mode button is pressed until a horizontal line appears on the
screen. This line is a profile of the brightness level of the image
versus the gun filament current (which would be zero at this moment).
- The filament current is slowly increased. The line profile will
develop peaks and valleys (which is indicative of the specimen
contrast) and will start to rise. It will go through a maximum, come
down and again rise to a maximum and is saturated at this value. The
filament current is set at the lowest value at which saturation occurs
(this is generally between 12 and 1'o clock positions on the filament
current knob).
- Change to Imaging Mode and start at low magnification. Focus and
solowly increase mag once an image is seen at low mag. Adjust
Contrast and Brightness correctly.
PROCEDURE for Cryo-SEM
The following steps are performed for preparing the sample.
- A drop of the specimen is placed on a sandwich holder.
- The sample assembly is plunged in a pool of melting freon or liquid
N2 with a plunger at high speed so that amorphous ice is formed.
- The sandwich is cryo-transferred into the SEM and held on a
cryo-holder within the SEM.
- The "frozen" sample is fractured by separating the sandwich or with the
help of special knife to expose an internal
surface and its embedded structures.
The sample is then imaged with SEM and snaps are taken.
- These photographs are then analysed by digital image analysing
modules associated with JSM-6400.
Sometimes, Freeze-fracture replication is useful, for both SEM and TEM:
Here is an extract from the "Materials and Methods" section of a paper
I wrote. Although the following paragraph is written for TEM, it is
also useful for SEM and Cryo-SEM:
Samples were observed for microstructure by transmission electron
microscopy using the
freeze-fracture cryo-vitrified replica technique as follows. Samples
for TEM were prepared by
high-pressure freezing using a Bal-Tec (Liechtenstein) Model HPM-010
with gold sample holders.
In this technique, a drop of sample is placed in a sample holder that
sandwiches the sample
between two gold sheets, one of which has a cup-shaped depression to
hold a few microliters of
the sample. The instrument rapidly freezes (vitrifies) the sample by
application of high
pressure followed by low-temeprature depressurization. The frozen
sample is transferred using a
cold-chain into a freeze fracture apparatus Bal-Tec (Liechtenstein)
Model BAF 060 Freeze
Fracture/Etch System, where the sample is kept on a liquid-nitrogen
cooled stage. An internal
surface of the sample is exposed by fracturing the sample by
separating the sandwich and assisted
by a built-in knife. The fracture surface, which has the internal
microstructure of the sample
exposed as topographic variations, is shadowed at 45 degree angle with
a platinum-carbon film
deposited by means of electron beam evaporation. Then a support layer
of carbon is evaporated
onto the shadowed layer, so as to give a replica that is brought up to
room temperature, washed
of residual sample, dried and observed by TEM. We used a FEI Technai
T-12 TEM operated at
120KeV.
Photographing the Image
Images can be photographed using the camera attached to the
microscope. Despite advances in digital photography, as of today, it
is only film, especially medium and large format film (120 size film and larger)that has the necessary resolution to capture the fine detail possible
with the SEM. The resolution applies to spatial resolution, the
number of pixels per unit length, as well as the bit-depth resolution.
Film used in the SEM approaches 4000 pixels per inch at 12 bits depth
(monochrome). The procedure is:
- Load the film - roll in the camera.
- Remove the shutter and get image in slow scan.
- Adjust optimum contrast and the brightness either by pressing "ACB"
button or manually.
- Press photo "RIGHT" or "LEFT" depending on which photo is to be
taken.
- With one roll, ten photos can be taken. After finishing ten
photographs, wind the roll fully till film forwarding tap gets
loosened.
- By releasing the key, open the camera and remove the roll from it.
Film Developing
There are three ways to process or develop a film roll:
- Give it to a professional film developer e.g. film studio or
CRL.
- Process it yourself in a tray.
- Process it yourself in a tank.
Processing it youself is best, as you can get immemdiate feedback that
your photographs are registered in a permanent form. In fact, I
highly
recommend that you keep your specimen in the SEM and develop and check
the film for OK micrographs before removing the specimen. This
ensures that your session has been successful. Success in this
instance is having good images of what you saw on the screen.
Tray processing: all steps must be done in total darkness
- Take approximately 500ml of the film developer in a tray.
- Take an equal amount of the fixer in another tray.
- Switch on the A.C. in the developing room and adjust it so that the
temperature in the room stabilizes at around 20 deg C. It takes about 20
minutes.
- Switch off all the lights in the dark room and the microscope room
and the sample preparation room.
- Cut the tape of the roll and unwind the film.
- Separate the film from the paper and hold the ends of it in two
hands.
- Set alarm for ten minutes.
- Start developing by dipping the film in the developer from one end
to the other in a see-saw action.
- Continue developing for ten minutes till alarm rings.
- Lift the film out of the developer.
- Drain excess developer into the tray
- Dip for 10 seconds in stop-bath (2% aqueous acetic acid)
- Wash the film with water
for approximately 5 minutes.
- Repeat the procedure as for developing, this time with fixing
solution.
- Continue fixing for 5 minutes.
- Now switch on the lights and wash the negative thoroughly under
water for about 10 minutes.
- Hang film to dry. Restore fixer and developer to respective
bottles.
- Note: Do not mix used fixer and developer with stock solution. Wash
and clean the trays and other apparatus.
Tank processing: first few steps must be done in total darkness;
thereafter, all steps can be done in room light. All chemical
handling is done in room light.
- Switch on the A.C. in the developing room and adjust it so that the
temperature in the room stabilizes at around 200c. It takes about 20
minutes.
- Switch off all the lights in the dark room and the microscope room
and the sample preparation room.
- Cut the tape of the roll and unwind the film.
- Separate the film from the paper and hold the ends of it in two
hands.
- Load the film into the spiral. (Learn how to do this by trying
with a used film in ordinary room light). Place the spiral into the
developing tank and close the lid. Now the room light can be turned
on.
- Set alarm for ten minutes.
- Start developing by dipping the spiral in the developer.
- Continue developing for ten minutes till alarm rings.
- Drain the developer from the tray into the deveoper bottle.
- Wash the film with water for approximately 2 minutes.
- Repeat the procedure as for developing, this time with fixing
solution.
- Continue it for 5 minutes.
- Wash the negative thoroughly under
water for about 10 minutes.
- Hang film to dry. Restore fixer and developer to respective
bottles.
- Note: Do not mix used fixer and developer with stock solution. Wash
and clean the trays and other apparatus.
Contact Printing
This is the simplest and most economical way to make prints from the developed roll for a
preliminary study of the results.
- Pour 500 ml of fixer and developer in two trays.
- Switch off all lights except red light.
- Place the negative to be printed in two aluminium slates with
printing paper beneath the negative to determine exact exposure
time(glossy side of the paper to face up).
- Cover up the film with a piece of paper so that only a short strip
of negative is exposed to light. Switch on the bulb for two seconds.
- Move paper covering the negative so that another short strip is
exposed and switch on the bulb for two seconds. Repeat same procedure
5 times.
- Now we have areas on the paper which are exposed to light 2,4,6,8
and 10 seconds.
- Remove test strip and develop in the developer tray for one and
half minutes.
- Wash the test strip in water and fix it in the fixer for about two
and half minutes. Again wash it with water and switch on the lights.
- Check the exposure time required from the test strip and switch
off all the lights except the red light.
- Now replace the negative between slates with fresh piece of
printing paper underneath it. Repeat the same steps as for the test
strip with exposure time found out from test strip.
Below is a list of items each operator should have. If possible, do
not share these items with others, as their quality may be suspect if
misused.
-
Developer
-
Fixer
-
Foreceps (blunt and pointed)
-
Scissors (Fiskars)
-
Knife/blade
-
Adhesive role
-
Screw driver(2/4)
-
Dropper
-
Syringes
-
Slides
-
Coverslips
-
Capillaries
-
Tissue paper
-
Silver paste
-
Paint brush (zero no.)
-
Fevicol
Contributors to this document are gratefully thanked. Principal
among them are: P. T. Joseph,
Rajgopal Iyer, Nikunj Parikh, Ashim Chowdhary, Mayur K. Temgire, Megha
Surve, and others.
Your comments are welcome.
Email: jb@iitb.ac.in