Protocol for starting the SAXS accessories before starting SAXS.
- Start
the chiller (check the chiller set temp is 25°C). The pump and compressor
are working properly
- Switch
on the power supply to the computer, Perkin
Elmer Cyclotron, Vaccum pump, Temperature
Controller Unit (TCU 50) and Water pump( Thermo
Haako). The switches are marked on the switch
board.
- Set
the temperature of the sample by selecting appropriate temperature in TCU
50 and press start button
Starting
the SAXS instrument
- Push
and hold the the Power On button on the Philips control panel of the SAXS instrument
till you see all the LEDs glow.
- Make
entry into the Log book by filling columns appropriately.
- Turn
the HT meter key to ON position
and wait till the display is 20 KV
10 mA and the OK led glows
- Slowly,
in steps of five units, increase the current and voltage alternately,
starting with current first, to 40
kV 40 mA using the + buttons of the HT
meter.
- Always
ensure that the OK LED is glowing. Incase the OK led is off please wait
till it glows again. Make the increments gradual and slow at a rate of one
unit per two seconds approximately.
- Open
the needle valve to release the vacuum slowly and steadily
- Open
the sample insertion lock and load the sample.
- Put
image plate in appropriate position by sliding it through the groove.
Ensure always that the cut side of the plate is always on top and the
X-Ray sensitive side (Blue side) faces the sample.
- Replace
the locker and turn it clock-wise. It is a pressure locking mechanism and
do not force any mechanical parts.
- Close
the needle valve and switch on the vacuum pump by pressing START/STOP press-button on the
vacuum pump display panel
- Program
the shutter
How
to program the shutter?
This is an example for programming shutter # 4 to open for 1
hour, 2 mins and 30 sec.
- Select
the shutter by pressing the shutter select push button switch until the
shutter # 4 is displayed.
- Press
the hrs/min/sec pushbutton
switch until the hrs digits are blinking in the display (Program mode).
- Press
the Time set switch + or pushbutton switch until the hrs digits read 01.
Each time this switch is pressed the time will advance or decreases one
increment. If the switch is held for at least 2 seconds, the time will
increase at a rapid rate for as long as the button is depressed.
- Press
the hrs/min/sec pushbutton
switch until the min digits are blinking in the display (Program mode)
- Press
the Time set switch + or pushbutton switch until the min digits read 02.
Each time this switch is pressed the time will advance or decreases one
increment. If the switch is held for at least 2 seconds, the time will
increase at a rapid rate for as long as the button is depressed.
- Press
the hrs/min/sec pushbutton
switch until the sec digits are blinking in the display (Program mode)
- Press
the Time set switch + or pushbutton switch until the min digits read 30.
Each time this switch is pressed the time will advance or decreases one
increment. If the switch is held for at least 2 seconds, the time will
increase at a rapid rate for as long as the button is depressed.
- Confirm
the display reads (shutter) # 4 = (status) CLOSED ( Timer ) 01 02 30
- Press
the hrs/min/sec pushbutton
switch one final time and see that all the digits are in a steady state. Press
the Assign pushbutton switch
for shutter # 4.
- Press
the Shutter open enable pushbutton switch and open # 4 switch. Simultaneously
and observe shutter # 4 OPEN and Timer starts counting down. Also observe
that the green light in the Assign pushbutton is illuminated.
- Read the Image Plate
Once the timer has counted down to
00-00-00, the shutter will close and after 2 seconds, the timer program will
return to its programmed setting (01-02-30).
If the
shutter is to be closed in the middle of the experiment due to some reason it
can be done using CLOSE pushbutton.
IF a shutter is closed while Assigned,
the timer will reset to the original setting. A shutter can be disassigned
by pressing ∞ and it will remain open indefinitely until manually
closed. If a shutter is disassigned and then
re-assigned, the timer will start at the original setting that was programmed
for that particular shutter.
Reading the image plate
- Start the cyclotron with the carrousel inside it.
Always ensure that the pins of the carrousel are normal to the center.
- Start the computer.
- Open OptiQuant software. Click instruments in the tool bar
at the top.
- The cyclotron gets initialized and then a virtual
grid of carrousel cut-open view is seen.
- Select the area to be scanned by selecting a mask
stored mask as per SAXS or WAXS experiment done.
- Shutter closes after counting down to 00-00-00. Then
release the vacuum by turning the needle valve slowly and steadily.
- After the vacuum reading is atmospheric, make the
room dark and remove the image plate. Note replace the lid on the SAXS
immediately after the plate is removed.
- Clamp the plate in the carrousel using the pins in
the dark room. Release the pins once the plate is in position
- Put the carrousel into the cyclotron and then close
its lid
- Click the start button on the OptiQuant software of the
initialized instrument window.
- The selected area is scanned and a tif
image of it is produced.
- Open the tif image in SaxsQuant software and
integrate the intensity data to get a I Vs q plot.
- Process the plot and save it as a .PDH file (Primary Data
Handling)
Instrument shut down.
- Release vacuum and remove sample holder from the SAXS
instrument.
- Replace the sample locker.
- Remove the image plate.
- Close the needle valve and start vacuum
- Decrease the voltage and current values of HT to 20
and 10 in steps of 5 units gradually, starting with voltage at a rate of 1
unit per 2 seconds.
- When the panel reads 20 kV and 10 mA
turn the HT key OFF.
- Make entry into the Log Book.
- Push the Power
Off button the panel to turn the power off.
- Shut down the computer and then switch off the
cyclotron.
- Stop vaccum and switch off
the TCU and ThermoHaako.
- Turn all the switches on the switch board off.
- Remove the sample from the sample holder and clean
thoroughly. Incase of liquid holders a prolonged cleaning session with
Millipore water is required.
- Turn the Chiller Off.
Check list before leaving the room
- Carrousel pins are in normal position.
- All instruments are off.
- All plates are on the eraser.
- Sample holders and other accessories back in the box.
- Sample removed from the sample holder.
- Sample holder properly cleaned.
- Log book entry done.
- Computer shutdown.